| Company | Stakpure GmbH | Stakpure GmbH | Stakpure GmbH | Stakpure GmbH | Stakpure GmbH |
| Item | OmniaPure xstouch | OmniaLabDS | OmniaLabED | OmniaLabED+ | OmniaTap xsbasic |
| Price | | | | | |
| Description | Type I ultrapure water system with intuitive touch screen (swipe and scroll functions), real-time TOC monitoring, integrated OptiFill touch dispenser, leakage sensor, and integrated pressure reducer. Available in UV-TOC and UV-TOC/UF variants. Part of the Blueline series. Supports bench, wall, and Type I ultrapure water system with intuitive touch screen (swipe and scroll functions), real-time TOC monitoring, integrated OptiFill touch dispenser, leakage sensor, and integrated pressure reducer. Available in UV-TOC and UV-TOC/UF variants. Part of the Blueline series. Supports bench, wall, and under-table installation.... Read More | The OmniaLabDS is a CLRW/CLSI-grade pure water system designed for feeding clinical analyzers, steam sterilizers, and washer-disinfectors. Available in 20, 40, and 60 l/h output models, it features a 100 L pressurized tank with quality recirculation, regenerative polishing cartridges, a 0.2 µm The OmniaLabDS is a CLRW/CLSI-grade pure water system designed for feeding clinical analyzers, steam sterilizers, and washer-disinfectors. Available in 20, 40, and 60 l/h output models, it features a 100 L pressurized tank with quality recirculation, regenerative polishing cartridges, a 0.2 µm sterile filter, UV disinfection, and an optional emergency supply. Complies with DIN EN 285 and EN ISO 15883 international water standards.... Read More | OmniaLabED is the efficient solution when high-quality pure water type II is required for the complete laboratory supply. The system meets international water standards ASTM, ISO 3696, and CLRW (CLSI), and uses continuously self-regenerating electro-deionization for maximized economy. It holds 100 OmniaLabED is the efficient solution when high-quality pure water type II is required for the complete laboratory supply. The system meets international water standards ASTM, ISO 3696, and CLRW (CLSI), and uses continuously self-regenerating electro-deionization for maximized economy. It holds 100 liters of pure water in a storage tank with quality recirculation to supply autoclaves, laboratory dishwashers, and ultrapure water systems. Available in two output models: OmniaLabED 20 (20 l/h) and OmniaLabED 40 (40 l/h).... Read More | The OmniaLabED+ is a large all-round ultrapure and pure water system designed for high-demand laboratories requiring both Type I ultrapure water and Type II pure water. Features continuously self-regenerating electro-deionization (EDI), a 100-liter recirculating storage tank with pressurized outlet,The OmniaLabED+ is a large all-round ultrapure and pure water system designed for high-demand laboratories requiring both Type I ultrapure water and Type II pure water. Features continuously self-regenerating electro-deionization (EDI), a 100-liter recirculating storage tank with pressurized outlet, and complies with ASTM, ISO 3696, and CLSI (CLRW) standards. Available in 20 L/h and 40 L/h variants with optional UV-TOC monitoring.... Read More | OmniaTap xsbasic is a compact allrounder system that produces both Type I ultrapure water and Type II pure water from a single unit connected directly to tap water. It features a OneHand (OptiFillbasic) dispenser for push-button dispensing of Type I ultrapure water, and supports adaptable pure waterOmniaTap xsbasic is a compact allrounder system that produces both Type I ultrapure water and Type II pure water from a single unit connected directly to tap water. It features a OneHand (OptiFillbasic) dispenser for push-button dispensing of Type I ultrapure water, and supports adaptable pure water tanks (7, 30, or 60 liters) for continuous Type II laboratory water withdrawal. Available in two configurations: standard (xSbasic5) and UV-enhanced (xSbasic5 UV) with lower TOC output.... Read More |
| Method | UV oxidation + TOC monitoring + mixed-bed ion exchange; UV-TOC/UF variant adds ultrafiltration (UF). Feed water prepared by ion exchange, reverse osmosis, electrodeionisation, or distillation. | RO cartridge ? stainless steel regenerative polishing cartridge ? 0.2 µm sterile filter capsule ? UV disinfection; sterile overflow and sterile vent filter + CO2 absorber included; optional UV tank disinfection and emergency supply | Reverse Osmosis (RO) + Electro-Deionization (EDI) | Reverse Osmosis (RO) + Electro-Deionization (EDI) + UV | Reverse osmosis with UV treatment; push-button digital dispenser control (OptiFillbasic OneHand dispenser); direct tap water feed connection |
| Output | Up to 2 L/min; individually adjustable dispensing volume 0.05 up to 25 liters | 20 l/h (OmniaLabDS20), 40 l/h (OmniaLabDS40), 60 l/h (OmniaLabDS60) at 15°C | 20–40 l/h (OmniaLabED 20: 20 l/h; OmniaLabED 40: 40 l/h) at 15 °C | 20 L/h or 40 L/h (Type II pure water); up to 2 L/min dispenser (Type I ultrapure water); 100 L/h at 2 bar pressurized tank outlet | 5 l/h pure water (Type II); up to 2 l/min ultrapure water (Type I) dispensing performance |
| Resistivity | 18.2 MO·cm at 25 °C (both UV-TOC and UV-TOC/UF variants) | 10 up to 1 MO·cm | 1–10 MO·cm (Conductivity: 0.1–1 µS/cm) | 18.2 MO·cm (Type I); 10 MO·cm down to 1 MO·cm (Type II conductivity 0.1–1 µS/cm) | 18.2 MO·cm at 25°C (Type I); 10–15 MO·cm at 25°C (Type II) |
| Dimensions | Standard unit (W x H x D): 240 x 530 x 435 mm; Production unit under table: 240 x 415 x 310 mm; Wall dispenser: 90 x 482 x 337 mm; Bench dispenser: 250 x 650 (750) x 380 mm | Tower: 511 x 1520 x 575 mm (W x H x D); Base cabinet (tank): 511 x 800 x 575 mm (W x H x D) | Tower (W x H x D): 511 x 1520 x 575 mm; Base cabinet/tank (W x H x D): 511 x 800 x 575 mm | Tower: 511 x 1520 x 575 mm (W x H x D); Base cabinet (tank): 511 x 800 x 575 mm (W x H x D) | 253 x 530 x 520 mm (W x H x D, with 7L docking tank) |
| Total Organic Carbon (TOC) | < 2 ppb (UV-TOC variant); < 5 ppb (UV-TOC/UF variant) | < 2 ppm (feed water requirement) | < 30 ppb (depending on feed water quality) | < 10 ppb (standard models); < 5 ppb (UV-TOC monitoring models) | < 10 ppb (Type I, standard model); < 5 ppb (Type I, UV model) |
| Get Quote | | | | | |