| Description | Compact pure water system with integrated reverse osmosis and deionization for Type II (ASTM II) pure water. Direct tap water connection, OptiFill touch dispenser standard, 6–20 L/h output, optional UV disinfection (254 nm), compatible storage tanks 10/30/60 L with recirculation. Applications: Compact pure water system with integrated reverse osmosis and deionization for Type II (ASTM II) pure water. Direct tap water connection, OptiFill touch dispenser standard, 6–20 L/h output, optional UV disinfection (254 nm), compatible storage tanks 10/30/60 L with recirculation. Applications: buffer and media preparation, feeding ultrapure water systems.... Read More | The OmniaLabUP is the reliable system for complete laboratory supply of high-quality Type II pure water. It holds 100 liters of pure water type II in a storage tank with quality recirculation and pressure outlet. OmniaLabUP supplies autoclaves, laboratory dishwashers, and ultrapure water systems. The OmniaLabUP is the reliable system for complete laboratory supply of high-quality Type II pure water. It holds 100 liters of pure water type II in a storage tank with quality recirculation and pressure outlet. OmniaLabUP supplies autoclaves, laboratory dishwashers, and ultrapure water systems. Available in 20 l/h and 40 l/h output models. Water produced conforms to ASTM, ISO 3696 and CLRW (CLSI) international water standards.... Read More | Compact, area-efficient ultrapure water (Type I) system with OptiFill basic OneHand dispenser. Delivers 18.2 MO·cm resistivity and TOC <10 ppb (standard) or <5 ppb (UV variant). Integrated leakage sensor, pressure reducer, and pre-treatment cartridge. Bench or wall mountable. Available as standard (Compact, area-efficient ultrapure water (Type I) system with OptiFill basic OneHand dispenser. Delivers 18.2 MO·cm resistivity and TOC <10 ppb (standard) or <5 ppb (UV variant). Integrated leakage sensor, pressure reducer, and pre-treatment cartridge. Bench or wall mountable. Available as standard (art. 18000004) and UV (art. 18000005) models. Typical applications: AAS, IC, ICP, buffers and media preparation (standard); ultra-trace analysis, ICP-MS, HPLC, TOC-analysis (UV).... Read More | Type I ultrapure water system with intuitive touch screen (swipe and scroll functions), real-time TOC monitoring, integrated OptiFill touch dispenser, leakage sensor, and integrated pressure reducer. Available in UV-TOC and UV-TOC/UF variants. Part of the Blueline series. Supports bench, wall, and Type I ultrapure water system with intuitive touch screen (swipe and scroll functions), real-time TOC monitoring, integrated OptiFill touch dispenser, leakage sensor, and integrated pressure reducer. Available in UV-TOC and UV-TOC/UF variants. Part of the Blueline series. Supports bench, wall, and under-table installation.... Read More | OmniaTap is the powerful all-rounder for simultaneous production of Type I ultrapure water and Type II pure water from a single unit. Available in 12 or 20 L/h configurations with optional UV-TOC monitoring and ultrafiltration. Connects directly to tap water. OptiFill touch dispenser with precise OmniaTap is the powerful all-rounder for simultaneous production of Type I ultrapure water and Type II pure water from a single unit. Available in 12 or 20 L/h configurations with optional UV-TOC monitoring and ultrafiltration. Connects directly to tap water. OptiFill touch dispenser with precise volume control. Optional 10, 30, or 60 L pure water storage tanks with recirculation. Suitable for AAS, IC, ICP, ICP-MS, HPLC, TOC analysis, buffers, media preparation, life science, and cell culture applications.... Read More |
| Method | Reverse osmosis (RO) + ion exchange deionization; optional UV disinfection at 254 nm; sterile filter capsule 0.2 µm (included) | Reverse osmosis + ion exchange polishing cartridge + 0.2 µm sterile filtration | Ion exchange with integrated pre-treatment cartridge; UV lamp (UV variant); recirculation function to prevent biofilm | UV oxidation + TOC monitoring + mixed-bed ion exchange; UV-TOC/UF variant adds ultrafiltration (UF). Feed water prepared by ion exchange, reverse osmosis, electrodeionisation, or distillation. | Reverse osmosis (RO) + ion exchange (ultrapure water cartridge) + sterile filtration 0.2 µm; UV lamp for TOC reduction (UV-TOC models); ultrafiltration/bio filter capsule (UV-TOC/UF model) |
| Dimensions | Tower without tank: 390 x 720 x 525 mm (W x H x D); Tower with 10-liter docking tank: 390 x 720 x 615 mm (W x H x D) | Tower (W x H x D): 511 x 1520 x 575 mm; Base cabinet tank (W x H x D): 511 x 800 x 575 mm | 240 x 530 x 435 mm (W x H x D) | Standard unit (W x H x D): 240 x 530 x 435 mm; Production unit under table: 240 x 415 x 310 mm; Wall dispenser: 90 x 482 x 337 mm; Bench dispenser: 250 x 650 (750) x 380 mm | Without tank (with OptiFill dispenser): 390 x 720 x 525 mm (W x H x D); With 10-L docking tank: 390 x 720 x 615 mm (W x H x D) |