Size(s) | 4, 5, 6-foot (1.2, 1.5, 1.8-meter) | Inquire | 3, 4, 5, 6-foot (0.9, 1.2, 1.5, 1.8-meter) | 3, 4, 6-foot (0.9, 1.2, 1.8-meter) | 6ft |
Description | The LabGard ES (Energy Saver) Model NU-677 Animal Handling Biosafety Cabinet (LFBSC) is a bench/table top model, with an integral castored support stand for operation as a console model. The LabGard model NU-677 utilizes an Energy Saver DC ECM motor optimally determined forward curved fan for each The LabGard ES (Energy Saver) Model NU-677 Animal Handling Biosafety Cabinet (LFBSC) is a bench/table top model, with an integral castored support stand for operation as a console model. The LabGard model NU-677 utilizes an Energy Saver DC ECM motor optimally determined forward curved fan for each model size/width to maximize both energy efficiency and filter loading capacity. The Energy Saver ECM motor is controlled via a solid-state DC motor controller that provides automatic compensation (constant volume control) for both filter loading and line voltage variances.
The Laminar Flow Biosafety Cabinet, (LFBSC) is a product resulting from the development of the ""laminar flow"" principle (see airflow schematic) and the application of environmental controls as required in the field of biological research or chemical containment. The LFBSC, when used with proper technique, is an effective laboratory aid in obtaining the optimum control over product quality while reducing the potential for exposure of both product and personnel to airborne biological or particulate chemical agents in low to moderate risk-hazard research and drug preparation or product operations, as prescribed by the Center for Disease Control (CDC) Atlanta, Georgia.
The NU-677 bench LFBSC is known as a Class II, Type A2 Biosafety Cabinet. This is possible since NuAire's cabinet conforms to the following requirements:
- Maintain a minimum average s inflow velocity of 100 ft/min (0.51 m/s) through the work access opening;
- Have HEPA/ULPA filtered downflow air that is a portion of the mixed downflow and inflow air from a common exhaust plenum;
- May exhaust HEPA/ULPA filtered air back into the laboratory or to the environment through an external exhaust system connected to the cabinet with a canopy connection; and
- Have all biologically contaminated ducts and plenums under negative pressure or surrounded by negative pressure ducts and plenums.
If working with volatile chemicals, the unit must be connected to an external exhaust system. Type A2 cabinets may be used for work with volatile chemicals if permitted by a chemical risk assessment.... Read More | Inquire | LabGard® ES (Energy Saver) AIR model NU-543 is defined as a Class II, Type A2 Biosafety Cabinet (BSC) providing the very best in personnel, product, and environmental protection. LabGard AIR features a single ultrahigh efficiency motor with a precision-engineered forward curve fan to increase LabGard® ES (Energy Saver) AIR model NU-543 is defined as a Class II, Type A2 Biosafety Cabinet (BSC) providing the very best in personnel, product, and environmental protection. LabGard AIR features a single ultrahigh efficiency motor with a precision-engineered forward curve fan to increase filter load capability and extend filter life. The motor is controlled to airflow setpoints by the FlowGard electronic control system with an integrated motor controller that provides automatic compensation for both filter loading and line voltage variances. Airflow velocity is monitored with the Intelliflow digital dual thermistor airflow sensor.
Highlights
Monolithic cabinet shell
Choose from 3 access opening heights [8, 10, 12-inches (203, 254, 305 mm)]
Up to 3 service valves each sidewall, smooth interior option
NSF approved cord pass through right side wall
NSF/ANSI 49, EN12469, UL-C, & CE Listed... Read More | The LabGard Containment Cabinet NU-813 is offered as a Class I Containment Cabinet equipped with a HEPA filter. In the Class I configuration, the NU-813 provides personnel protection for handling low risk aerosols and particulate matter. Room air enters the Cabinet, flowing past the operator into The LabGard Containment Cabinet NU-813 is offered as a Class I Containment Cabinet equipped with a HEPA filter. In the Class I configuration, the NU-813 provides personnel protection for handling low risk aerosols and particulate matter. Room air enters the Cabinet, flowing past the operator into the work access opening providing an air barrier which minimizes contaminants generated within the cabinet work zone from migrating into the room. The air flows through the cabinet carrying contaminants with it, exiting through a HEPA filter as a Class I Cabinet.... Read More | The LabGard NU-610 Dual Access Laminar Flow, Class II, Type A2 Biological Safety Cabinet (LFBSC) is a console model, a product resulting from the development of the "laminar flow" principle and the application of environmental controls as required in the field of biological research or chemical The LabGard NU-610 Dual Access Laminar Flow, Class II, Type A2 Biological Safety Cabinet (LFBSC) is a console model, a product resulting from the development of the "laminar flow" principle and the application of environmental controls as required in the field of biological research or chemical containment. The LFBSC, when used with proper technique, is an effective laboratory aid in obtaining the optimum control over product quality while reducing the potential for exposure of both product and personnel to airborne biological or particulate chemical agents in low to moderate risk-hazard research and drug preparation or product operations, as prescribed by the Center for Disease Control (CDC) Atlanta, Georgia.... Read More |