Description | The Nicomp dynamic light scattering (DLS) system measures size and zeta potential of nano sized particles from 0.3 nm to 10 µm. The backbone of the Nicomp DLS system is a proprietary high resolution multimodal deconvolution algorithm that can resolve close bimodal distributions. More The Nicomp dynamic light scattering (DLS) system measures size and zeta potential of nano sized particles from 0.3 nm to 10 µm. The backbone of the Nicomp DLS system is a proprietary high resolution multimodal deconvolution algorithm that can resolve close bimodal distributions. More importantly, it can separate the native peak of a distribution from the aggregate tail allowing researchers to determine the size of the native population. This is a critical parameter when trying to develop materials where the native population is directly responsible for the final properties of a product.
The Nicomp DLS instrumentation is engineered using a modular design approach. This allows for semi customization of the instruments based on the customers application needs.
Standard Nicomp Features
- Nicomp high resolution multi modal algorithm.
- Baseline adjust for automatic aggregate compensation.
- Multiple sample cells including a 1 cm cuvette and 6mm culture tubes that are low volume and can be centrifuged to remove contaminates.
- Time history plot provides a visual display of when adequate scattering data is collected to ensure statistically valid results.
- Vesicle Weighting – provides a light scattering correction for analysis of unilamellar vesicles.
Optional Features
- Patented Autodilution – eliminates the need for manual dilution of concentrated samples up to 40% solids. Provides more precise dilutions to the optimal measuring concentration resulting in higher reproducibility. It is the enabling technology for automation; batch autosampler and online measurements.
- Multi-Angle Goniometer – utilizes a precision stepper motor coupled with an optical fiber to provide multi angle sizing from approximately 12 – 175 degrees. Ideal for advanced users with complex distributions in the Mie regime that are looking to focus in on parts of their distribution.
- High Power Laser Diodes – 15 – 100 mW red, 20 – 100 mW green provide adequate scattering for ultra-small particles below 20 nm and low scattering materials avoiding without the need to compromise the measurement by over concentrating the sample which leads to multiple scattering and particle interaction effects.
- Avalanche Photo Diode Detector – provides a high sensitivity detector for low concentration and low scattering samples. Up to 7× gain over a standard photomultiplier tube and when coupled with a higher power laser diode enables accurate sizing information for sub 1 nm particles.
- Low Volume Cells – index matched sample cells for analysis of samples volumes down to 10 microliters. Ideal for expensive protein and macromolecules used in life science applications.
- Autosampler/Online – by incorporating the auto dilution feature in the Nicomp which can handle samples with up to 40% solids, the measurement can be extended into a batch auto sampler platform for automated QC analysis of up to 96 samples
... Read More | The AccuSizer is a sophisticated high-resolution particle counter and particle size analyzer that uses the technique of singal particle optical sizing (SPOS). The major benefits of this technique include:
- High Accuracy: Each particle is measured directly
- Concentratation: accurate particles/mL
The AccuSizer is a sophisticated high-resolution particle counter and particle size analyzer that uses the technique of singal particle optical sizing (SPOS). The major benefits of this technique include:
- High Accuracy: Each particle is measured directly
- Concentratation: accurate particles/mL results
- Resolution: no distribution broadening, split multiple peaks to baseline, detect small concentration of tails
AccuSizer AD System - Patented Autodilution
- The most popular for general particle size analysis applications
- Performs exponential autodilution while providing accurate concentration information
The autodilution system includes an input, output, dilution chamber, sample material, and diluent. A known quantity of sample material is injected into the dilution chamber. Filtered diluent flows into the dilution chamber, mixes with the sample material, and continuously dilutes the initial concentration. The diluted sample then flows out of the dilution chamber and through the sensor for particle size and count analysis.... Read More | The AccuSizer is a sophisticated high-resolution particle counter and particle size analyzer that uses the technique of singal particle optical sizing (SPOS). The major benefits of this technique include:
- High Accuracy: Each particle is measured directly
- Concentratation: accurate particles/mL
The AccuSizer is a sophisticated high-resolution particle counter and particle size analyzer that uses the technique of singal particle optical sizing (SPOS). The major benefits of this technique include:
- High Accuracy: Each particle is measured directly
- Concentratation: accurate particles/mL results
- Resolution: no distribution broadening, split multiple peaks to baseline, detect small concentration of tails
AccuSizer FX-Nano AD System
- Uses LE400 plus FX-Nano sensors with AD-style exponential autodilution down to 150 nm
- Can analyze high concentration samples up to 106 particles/mL
- Ideal for filter testing and high concentration protein samples
... Read More | A comprehensive portfolio of high accuracy and resolution liquid particle counters that determine particle size and concentration of your suspensions.
Features:
- Uses single particle optical sizing (SPOS) technology to provide extended dynamic size and concentration ranges
-
Can be used as a liquid
A comprehensive portfolio of high accuracy and resolution liquid particle counters that determine particle size and concentration of your suspensions.
Features:
- Uses single particle optical sizing (SPOS) technology to provide extended dynamic size and concentration ranges
-
Can be used as a liquid particle counter for contamination applications or as sophisticated particle size analyzer
-
Highest sensitivity technology available to determine tails of distribution such as large particle counts (LPC) in CMP slurries, protein aggregates, or tails of emulsions (PFAT5)
-
Fast, multi-channel pulse height analyzers that can report results in up to 1024 size bins
-
Several fluidics configurations are available to deliver the sample to the sensor directly or with auto dilution to avoid coincidence errors
... Read More | Nanoparticles are now standard materials for many industrial products. Like most other processes, the production of nano-particles needs to be measured in order to be controlled. Dynamic light scattering (DLS) measurements have traditionally been performed using laboratory instruments like the Nanoparticles are now standard materials for many industrial products. Like most other processes, the production of nano-particles needs to be measured in order to be controlled. Dynamic light scattering (DLS) measurements have traditionally been performed using laboratory instruments like the Entegris Nicomp system, but the Entegris particle sizing team also has many years of online experience. The Mini DLS builds on these years of experience to develop a flexible and elegant solution that can be adapted to a wide range of nanoparticle manufacturing processes including milling, homogenizing, or self-assembly.
On-line particle size analyzer for nanoparticles and emulsions from 20 nm to 2 µm
- Flexible protocols to optimize measurement to application based on dynamic light scattering (DLS)
- Automated sampling and measurement to eliminate operator and lab resources
- Continuous data to track changes in particle size distribution
- Easy data transfer to process monitoring control software
... Read More |
Detector(s) | 15 – 100 mW red laser diodes, 20 – 100 mW green laser diodes | Extinction, Scattering, Collimated, and/or Focused (sensor dependent) | Focused beam, scatter 0.15-10+ µm at high concentration | Extinction and Scattering | Inquire |
Applications | Materials Testing, Pharmaceutical Lab | Materials Testing, Pharmaceutical Lab | Materials Testing, Pharmaceutical Lab | Inquire | Submicron suspensions and emulsions |