Tech Compare: Single vs. Double Beam Spectrophotometers

 Tech Compare: Single vs. Double Beam Spectrophotometers

Spectrophotometers measure light intensity as a function of wavelength and are commonly used to measure the concentration of a compound in an aqueous solution. Depending on the type of spectrophotometer, different wavelengths of light can be analyzed.

While wavelength range and detection limits are undoubtedly the most important factors to consider when selecting a proper spectrophotometer, there are other important criteria such as:

  • The density, shape, or size of the product you wish to measure
  • Analytical working range
  • Sample throughput
  • Data quality
  • Customizable and/or pre-configured method options
  • Measurement time1

There are two main types of spectrophotometers: single beam and double beam. As their names indicate, the major difference between the two instruments is the number of beams of light used in analysis.

Single Beam

In a single beam spectrophotometer, all the light waves coming from the light source pass through the sample as one beam. The instrument is standardized by placing a reference in the sample holder, and the resulting value is subtracted from subsequent sample measurements to remove effects from the solvent and the cell. The sensitivity of detection of the light beam after it passes through the sample is high since it uses a non-split light beam—therefore, high energy exists throughout.

Single-beam UV-Vis spectrophotometers can measure in the wavelength range 190 to 750 nm, although some go up to 1100. The UV region, considered to be any wavelength less than 340 nm, is useful for measuring nucleic acids, purified proteins and other organic molecules, for example.

Double Beam

Conversely, light coming from the source of a double beam spectrophotometer splits into two beams, with only one passing through the sample, and the other only passing through the reference. Each monochromatic beam in turn is split into two equal intensity beams by a half-mirrored device. One beam, the sample beam, passes through a cuvette containing a solution of the compound being studied in a transparent solvent. The other beam, the reference, passes through an identical cuvette containing only the solvent. The intensities of these light beams are then measured by electronic detectors and compared. Double beam spectrophotometers typically offer a measurement range of 190 to 1100 nm.

While not considered an additional type of spectrophotometer, split-beam instruments are similar to double-beam spectrophotometers. However, instead, they rely on a beam-splitter that rapidly alternates the light path between the reference and the sample while using one detector.

Advantages and Disadvantages

While double beam spectrophotometers are the more popular choice in today’s modern laboratories, there are still some advantages to single beam spectrophotometers. The cost-effectiveness of single beam spectrometers is the largest benefit, as well as a smaller footprint compared with other spectrophotometers. Single beam spectrophotometers also exhibit high sensitivity of detection due to high energy throughput from the non-splitting of the source beam.

On the other hand, if an application calls for stability, a single-beam spectrophotometer may not be the best option as the instrument does not compensate for disturbances like electronic circuit fluctuations, voltage fluctuations, mechanical component instability, or drift in the energy of light sources. Such drifts cause abnormal fluctuations in results.2

A double beam instrument, meanwhile, permits compensation for source power fluctuations, greatly improving signal-to-noise and extension to dilute solution samples and measurements with gases.

In general, double beam spectrophotometers operate faster and provide more reproducible results because they perform an automatic correction for the loss of light intensity as the beam passes through the sample and reference solution. Additionally, there is little or no lamp warm-up time, which not only helps improve the throughput of results but also conserves lamp life.

Double beam spectrophotometers are significantly more expensive than single-beam versions, though the lifetime cost of the instrument can be reduced when conversation of lamp life is taken into account.

Cuvettes

Regardless of if there’s one beam or two, cuvettes are critical to successful spectrophotometer analysis. There are many types of cuvettes available, even when narrowing down the selection to only those used for absorbance measurements in UV-Vis spectrophotometry—meaning the cuvettes must be transparent to the wavelengths used during chemical analysis.

Arguably the most important factor when selecting a UV-Vis cuvette is material. The two most common cuvette materials are glass and quartz.

Optical glass

Optical glass material cuvettes are a good choice on a tight budget. Here, the desired wavelength range is visible spectral (340 to 2500 nm). The material has a good transmission rate of >80% at 350 nm. Most measurements will fall into this area and it’s not necessarily required to have the extra UV quartz material range (190 to 340 nm) for select applications.3

UV quartz

Another option is UV quartz, which is a step above glass material. UV quartz material has a measurement transmission range of 190 to 2500 nm.3 Quartz and other UV-compatible plastics generally feature better absorption across a broader wavelength range—a requirement for measuring samples in spectrometers with UV-Vis light sources.4

IR quartz

The transmission range for IR quartz cuvettes is 230 to 3500 nm, allowing for some of the UV spectrum and an extended range in the infrared as well. An IR quartz cuvette may be required to offer greater advantages in IR applications where the highest sample purity is needed.4

Plastic

When high speed is the most important research criteria, plastic cuvettes are a suited choice. Available with a usable range of 380 to 780 nm (the visible spectrum range), plastic cuvettes are disposable after use, which comes with the benefit of zero chance of cross-contamination. Inexpensive to manufacture and purchase, disposable plastic cuvettes have been widely used in research labs where the beam light is not high enough to have an effect on the absorption consistency and tolerance value.3

References

  1. "Spectroscopy Applications Selection Guide," Cole-Parmer. https://www.coleparmer.com/tech-article/spectroscopy-selection-guide
  2. "Comparison between Single Beam and Double Beam Atomic Absorption Spectrometer Systems," Article by Deepak Bhanot, Lab-Training.com (2013). https://lab-training.com/comparison-between-single-beam-and-double-beam-atomic-absorption-spectrometer-systems/
  3. "Explained: UV vis Spectrophotometer and Fluorescence Cuvettes," CotsLab. https://cotslab.com/guide-uv-vis-spectrophotometer-and-fluorescence-cuvettes/#types-of-cuvette-materials
  4. "Guide to Cuvettes," SpectrEcology (2022). https://spectrecology.com/blog/guide-to-cuvettes/

 