The methodology described within United States EPA Method 8270 for the analysis of semivolatile organic compounds by GC/MS is applicable to other regions worldwide. The method presents several analytical challenges due to the requirement for simultaneous measurement of diverse classes of compounds over a wide concentration range. The 5977A and B Inert Plus GC/MSD meet the performance requirements of Method 8270D/E over a working range of 0.2–160 ppm using a single injection (one calibration). Initial calibration results are indicative of how long the continuing calibration will last and, thus, the length of time samples may be analyzed without intervention by the operator, which lowers the cost of operation. A wider calibration range and low compound %RSDs were achieved with this method.

Download Now

    Please fill out the information below to request information on the products you selected. If you are already registered, or have requested a quote before, and this information will be filled out for you.
  • Please enter your first and last name.

  • Please enter the name of the company or institution you work at.

  • Please enter your mailing address.

  • Please enter the e-mail address you would like to be contacted at.

  • Please enter the phone number you would like to be contacted at.

  • By submitting this form, you consent to Labcompare storing your email address and contact information and transmitting your contact information to the Application note sponsor. You may request to be removed at any time.