Size(s) | Inquire | 4, 5, 6-foot (1.2, 1.5, 1.8-meter) | 3, 4, 5, 6-foot (0.9, 1.2, 1.5, 1.8-meter) | 3, 4, 5, 6-foot (0.9, 1.2, 1.5, 1.8-meter) | 3, 4, 6-foot (0.9, 1.2, 1.8-meter) |
Description | The LabGard® ES LP (Low Profile) NU-640 Class II, Type A2 Animal Handling Biosafety Cabinet provides the largest work access opening with the smallest footprint while offering enough versatility to assist with all facets in the vivarium industry. 105 fpm [0.53 m/s] in-flow combined with 60 fpm [ The LabGard® ES LP (Low Profile) NU-640 Class II, Type A2 Animal Handling Biosafety Cabinet provides the largest work access opening with the smallest footprint while offering enough versatility to assist with all facets in the vivarium industry. 105 fpm [0.53 m/s] in-flow combined with 60 fpm [0.30 m/s] true laminar down-flow creates the dynamic air barrier. The nominal 6 foot [1.8 m] wide cabinet allows the ability to mix and match an assortment of configurations. Combine Refuse Disposal with Dirty Cage Collection to create your perfect animal waste station. The NU-640 also fits through standard doorways. ... Read More | The LabGard ES (Energy Saver) CYTO Model NU-581 Laminar Flow Biosafety Cabinet (LFBSC) is a console model. The LabGard Model NU-581 utilizes an Energy Saver DC ECM Motor optimally determined forward curved fan for each model size/width to maximize both energy efficiency and filter loading capacity. The LabGard ES (Energy Saver) CYTO Model NU-581 Laminar Flow Biosafety Cabinet (LFBSC) is a console model. The LabGard Model NU-581 utilizes an Energy Saver DC ECM Motor optimally determined forward curved fan for each model size/width to maximize both energy efficiency and filter loading capacity. The Energy Saver ECM Motor is controlled to airflow setpoints via a solid-state DC motor controller with digital dual thermistor airflow sensors that provide an automatic compensation (constant volume control) for both filter loading and line voltage variances.
The Laminar Flow Biosafety Cabinet, (LFBSC) is a product resulting from the development of the ""laminar flow"" principle (see Airflow Schematic) and the application of environmental controls as required in the field of biological research or chemical (hazardous drugs)containment. The LFBSC, when used with proper technique, is an effective primary engineering control in obtaining the optimum product quality while reducing the potential for exposure of both product and personnel to airborne biological or particulate chemical (hazardous drugs) agents in low to moderate risk-hazard research and drug preparation or product operations, as prescribed by the Center for Disease Control (CDC) Atlanta, Georgia.
The NU-581 Console LFBSC meets the requirements of a Class II, Type A2 since the cabinet conforms to the following requirements:
- Maintains a minimum inflow velocity of 100 LFPM (.51mps) through the work access opening
- Has HEPA filtered downflow air that is mixed with the inflow air from a common exhaust plenum
- Discharges a percentage of air to the outside atmosphere after HEPA filtration
- Has all biologically contaminated ducts and plenums under negative pressure or surrounded by negative pressure
Cabinets used for work with minute quantities of non-flammable or explosive volatile toxic chemicals and tracer amounts of radionuclides required as an adjunct to microbiological studies must be exhausted through properly functioning exhaust canopies.... Read More | The LabGard® ES NU-540 Class II, Type A2 Laminar Flow Biological Safety Cabinet (LFBSC) is an effective primary engineering control in obtaining the optimum product quality while reducing the potential for exposure of both product and personnel to airborne biological or particulate chemical agents The LabGard® ES NU-540 Class II, Type A2 Laminar Flow Biological Safety Cabinet (LFBSC) is an effective primary engineering control in obtaining the optimum product quality while reducing the potential for exposure of both product and personnel to airborne biological or particulate chemical agents in low to moderate risk-hazard research and drug preparation.The NU-540 offers economical operation through the use of an ultra high efficiency motor to maximize filter life and monolithic silicone-free construction for long product life.... Read More | LabGard® ES (Energy Saver) AIR model NU-543 is defined as a Class II, Type A2 Biosafety Cabinet (BSC) providing the very best in personnel, product, and environmental protection. LabGard AIR features a single ultrahigh efficiency motor with a precision-engineered forward curve fan to increase LabGard® ES (Energy Saver) AIR model NU-543 is defined as a Class II, Type A2 Biosafety Cabinet (BSC) providing the very best in personnel, product, and environmental protection. LabGard AIR features a single ultrahigh efficiency motor with a precision-engineered forward curve fan to increase filter load capability and extend filter life. The motor is controlled to airflow setpoints by the FlowGard electronic control system with an integrated motor controller that provides automatic compensation for both filter loading and line voltage variances. Airflow velocity is monitored with the Intelliflow digital dual thermistor airflow sensor.
Highlights
Monolithic cabinet shell
Choose from 3 access opening heights [8, 10, 12-inches (203, 254, 305 mm)]
Up to 3 service valves each sidewall, smooth interior option
NSF approved cord pass through right side wall
NSF/ANSI 49, EN12469, UL-C, & CE Listed... Read More | The LabGard Containment Cabinet NU-813 is offered as a Class I Containment Cabinet equipped with a HEPA filter. In the Class I configuration, the NU-813 provides personnel protection for handling low risk aerosols and particulate matter. Room air enters the Cabinet, flowing past the operator into The LabGard Containment Cabinet NU-813 is offered as a Class I Containment Cabinet equipped with a HEPA filter. In the Class I configuration, the NU-813 provides personnel protection for handling low risk aerosols and particulate matter. Room air enters the Cabinet, flowing past the operator into the work access opening providing an air barrier which minimizes contaminants generated within the cabinet work zone from migrating into the room. The air flows through the cabinet carrying contaminants with it, exiting through a HEPA filter as a Class I Cabinet.... Read More |