Description | Method scouting is a process where different mobile phases and column stationary phases are evaluated to determine the best set of separation conditions for a particular sample. Manual solvent preparation, column switching, and method programming are time-consuming steps, so the Nexera method Method scouting is a process where different mobile phases and column stationary phases are evaluated to determine the best set of separation conditions for a particular sample. Manual solvent preparation, column switching, and method programming are time-consuming steps, so the Nexera method scouting system was developed to streamline the entire process for more efficient method development.
The system is based around 2 LC-30AD pumps, each with a quaternary valve, which allows the analyst to run binary gradients with 16 different solvent pairs. A high-pressure column selection valve holds from 1 to 6 columns, resulting in up to 96 unique separation conditions per sample. Ultra-high speed conditions can be easily transferred to conventional conditions using a transfer program. In addition, scouting is performed automatically when the mobile phase is switched.
The Nexera Method Scouting Solution software has intuitive, user-friendly screens to set the pump and column valve positions, gradient program steps, and it automatically builds the sequence table in seconds. Method Scouting Solution provides the data required for analysis via a graphical user interface, as well as automated controls conforming to laboratory operations configured in the system settings.
... Read More | The CBM-40 is a small-profile unit that stacks neatly with main system components. Other LC modules connect via fiber optic cables for easy "plug and play" system configuration. An internal web server offers convenient control and monitoring of the UHPLC system through a web browser, LabSolutions The CBM-40 is a small-profile unit that stacks neatly with main system components. Other LC modules connect via fiber optic cables for easy "plug and play" system configuration. An internal web server offers convenient control and monitoring of the UHPLC system through a web browser, LabSolutions CDS, or other third-party software.
Features:
- Connection via Ethernet
- Offers 4 Fiber Optic ports for controlling other system components
... Read More | This system can switch automatically between SFC analysis and UHPLC analysis and make measurements on a single sample in each separation mode. It enhances user-friendliness and efficiency by allowing the investigation of separation conditions and performing reverse-phase high-speed analysis in a This system can switch automatically between SFC analysis and UHPLC analysis and make measurements on a single sample in each separation mode. It enhances user-friendliness and efficiency by allowing the investigation of separation conditions and performing reverse-phase high-speed analysis in a single system.
Key benefits:
- Enables both UHPLC and SFC analysis using a single system
- Reduces laboratory space requirements and equipment costs
- Existing UHPLC can be upgraded to an SFC/LC Switching System
- Method Scouting Solution software automates and manages method development analyses; switching from UHPLC to SFC conditions is performed without user intervention
- Reliable analysis even when switching between SFC and UHPLC modes
... Read More | Nexera XR is perfect for fast separations using the popular SPP column phases for high efficiency and lower back pressure. This system offers a rugged and forgiving UHPLC workflow, with five to ten-minute separations. With Analytical Intelligence functions, the system makes lab management simple, Nexera XR is perfect for fast separations using the popular SPP column phases for high efficiency and lower back pressure. This system offers a rugged and forgiving UHPLC workflow, with five to ten-minute separations. With Analytical Intelligence functions, the system makes lab management simple, enabling users to easily review instrument status, optimize resource allocation, and achieve higher throughput.
Key features and benefits include:
- High efficiency chromatography with 70 MPa (~10,000 psi) pressure tolerance and LPGE or HPGE solvent delivery
- Ultra-fast injection cycles (7 sec) with ultralow carryover
- High capacity sample handling
- Optional dual-injection system
- Outstanding baseline stability and linearity (2.5 AU), even under fluctuating temperatures
- Plate changers enable non-stop temperature-controlled analysis of thousands of samples
- Allows remote monitoring in real time directly from a web browser or smart device
- Touch-panel interface allows the user to control the instrument and carry out analysis preparation directly without the need for a PC
- Analytical intelligence functions, including smart flow control, automatic startup, auto-recovery and auto-diagnostics, and real-time mobile phase monitoring
- i-PeakFinder tool – can automatically recognize peaks even in complex chromatograms with high noise, and accounts for baseline drift for higher integration accuracy
- i-PDeA II peak deconvolution function – can quantify peaks that cannot be completely separated by a column
... Read More | Enables Both HPLC and UHPLC Analysis
The Nexera-i MT is a new addition to the i-Series lineup. It enables both conventional analysis via its high compatibility with existing LC systems and rapid analysis via high-speed methods. In pharmaceutical development, a single integrated system allows Enables Both HPLC and UHPLC Analysis
The Nexera-i MT is a new addition to the i-Series lineup. It enables both conventional analysis via its high compatibility with existing LC systems and rapid analysis via high-speed methods. In pharmaceutical development, a single integrated system allows process synthesis screening analysis for synthetic substances in the UHPLC flow line, and impurity content identification in the HPLC flow line.
Analysis results from LC systems in the same laboratory may differ, even with the same method, due to differences in the volumes of these systems. The Nexera-i MT demonstrates outstanding reproducibility between various LC systems with different system volumes, simplifying method transfer between instruments.
Further more, with the Nexera-i MT the process of migrating from UHPLC to HPLC or from HPLC to UHPLC can be accomplished using a single system. The Nexera-i MT is based on the same user-friendly i-Series, but maximizes performance and reliability for migrating and transferring customer methods.... Read More |