How to Avoid Common Sources of Error in Laboratory Mercury Testing

 How to Avoid Common Sources of Error in Laboratory Mercury Testing

by Michael Halvorson, PhD, Business Development Manager, AGS Scientific

Accurate mercury testing is essential for laboratories involved in environmental monitoring, industrial quality control, food safety, and regulatory compliance. Mercury is a toxic element that can pose risks to human health and the environment, making reliable measurements important for assessing contamination, evaluating potential exposure, and meeting applicable testing requirements. 

Even small analytical errors can affect reported mercury concentrations, particularly when laboratories measure trace amounts in wastewater, soil, food products, and other materials. Contamination, improper sample handling, instrument performance issues, and analytical interferences can produce inaccurate results that affect scientific conclusions and regulatory decisions. 

Understanding the most common sources of error in laboratory mercury testing can help scientists, laboratory managers, and researchers improve analytical reliability. By following established procedures throughout sample collection, preparation, analysis, and reporting, laboratories can reduce uncertainty and produce more consistent results. 

Sample Contamination During Collection and Preparation 

Sample contamination is a common source of error in mercury testing, particularly when analyzing samples containing extremely low concentrations. Even small amounts of mercury introduced during collection or preparation can produce artificially elevated measurements. 

Contamination can originate from sampling containers, laboratory glassware, reagents, equipment, and the surrounding environment. For example, containers that have not been properly cleaned may contain residual mercury from previous use. Similarly, reagents containing trace impurities can introduce mercury during sample preparation. 

Cross-contamination presents another concern when laboratories process multiple samples using shared equipment. Without appropriate cleaning between samples, mercury residues may transfer from one sample to another, affecting analytical accuracy. 

Laboratories can reduce contamination risks by using containers and equipment that meet the requirements of the applicable analytical method. High-purity reagents, established cleaning procedures, and controlled laboratory environments can further reduce the risk of introducing mercury contamination into samples. 

Field blanks and laboratory blanks also play an important role in identifying contamination. By analyzing appropriate blanks alongside actual samples, laboratory personnel can identify potential contamination introduced during collection, transportation, preparation, or analysis. 

Improper Sample Handling and Storage 

Mercury concentrations can change between sample collection and laboratory analysis when samples are handled or stored improperly. These changes may produce results that do not accurately represent the original sample. 

One concern is mercury loss through volatilization, particularly when samples contain elemental mercury or other volatile mercury species. Mercury may also adsorb onto container surfaces or undergo chemical transformations that affect its recovery during analysis. 

Improper storage conditions can increase these risks. Inadequate preservation, unsuitable containers, or excessive storage periods may contribute to changes in mercury concentration or chemical form. The appropriate handling procedures depend on the sample matrix, mercury species being measured, and analytical method. 

For example, certain aqueous mercury testing methods require chemical preservation to maintain sample integrity. However, preservation requirements for water samples may differ from those used for soil, biological materials, or other sample types. 

Laboratories should follow method-specific requirements for container selection, preservation, transportation, storage temperatures, and holding times. Maintaining complete sample documentation, including collection dates, preservation procedures, and storage conditions, also helps laboratory personnel identify potential handling errors before reporting results. 

Errors in Sample Preparation and Digestion 

Sample preparation is a critical stage of mercury testing because many analytical methods require mercury to be converted into a form suitable for measurement. Errors during preparation can affect mercury recovery and introduce uncertainty into the final results. 

Incomplete digestion or oxidation is one potential source of error. When sample digestion or oxidation is incomplete, some mercury may remain unavailable for measurement, resulting in incomplete recovery. This can produce results that underestimate the actual concentration. 

Preparation errors may also involve incorrect reagent concentrations, inaccurate sample volumes, unsuitable temperatures, or inconsistent processing times. These problems can affect the chemical reactions required for reliable mercury measurement. 

Certain preparation procedures also introduce opportunities for contamination or mercury loss. For example, excessive heating or inappropriate handling of open containers may increase the potential for losing volatile mercury species. 

Not every mercury testing method requires the same preparation process. Some direct mercury analysis techniques use thermal decomposition, followed by mercury collection and detection, to measure mercury in suitable samples without conventional acid digestion. Laboratories should therefore follow the preparation requirements established for their specific analytical method. 

Using validated procedures, properly maintained equipment, and appropriate quality control samples can help laboratories identify preparation-related errors. Recovery checks are particularly useful for determining whether the preparation process is producing acceptable results. 

Instrument Calibration and Performance Issues 

Instrument calibration is essential for establishing a reliable relationship between analytical signals and mercury concentrations. Improper calibration can introduce systematic errors that affect multiple samples, even when collection and preparation procedures are performed correctly. 

Calibration errors can occur when standards are improperly prepared, contaminated, unstable, or unsuitable for the intended measurement range. These problems can affect the accuracy of the calibration curve and lead to incorrect mercury measurements. 

Instrument drift is another concern. Changes in detector response, operating conditions, or instrument components may cause measurements to shift over time. Laboratories may also encounter unstable baselines, sample carryover, or contamination within the analytical system. 

Different mercury measurement techniques have distinct operating requirements. Cold vapor atomic absorption spectroscopy and cold vapor atomic fluorescence spectroscopy, for example, depend on appropriate chemical conditions and instrument performance to produce reliable measurements. 

When evaluating ⁠mercury analyzers, laboratories should consider the analytical technique, required detection limits, sample matrices, and applicable testing methods. Selecting equipment suited to the intended application is an important part of maintaining reliable analytical performance. 

Routine calibration verification can help laboratories detect changes in instrument response before they affect additional results. Preventive maintenance, performance checks, and established troubleshooting procedures also support consistent operation. 

When calibration verification results fall outside method-specific acceptance criteria, laboratories should investigate the problem, take corrective action, and determine whether previously analyzed samples require reanalysis. 

Matrix Effects and Analytical Interferences 

The chemical composition of a sample can influence mercury measurement accuracy. These effects, commonly referred to as matrix effects, occur when substances present in the sample interfere with mercury recovery, chemical reactions, or instrumental detection. 

Different sample types present different analytical challenges. Wastewater may contain dissolved solids, organic compounds, and industrial chemicals, while soil and sediment samples may contain minerals and complex organic materials. Food products and biological samples can introduce additional substances that affect preparation and analysis. 

High concentrations of organic matter may interfere with oxidation or digestion procedures if the selected preparation method is inadequate. Certain chemical compounds may also affect mercury reduction reactions or instrumental measurements, depending on the analytical technique. 

These interferences can produce artificially high or low results, reduce analytical sensitivity, or contribute to inconsistent measurements between samples. 

Laboratories should evaluate potential matrix effects when selecting and validating mercury testing procedures. Matrix spikes can help laboratories evaluate analytical recovery by measuring how much of a known mercury addition is recovered from a sample. 

Other approaches may include sample dilution, additional preparation steps, interference control procedures, or alternative analytical methods. However, these adjustments must remain consistent with the applicable testing method and its performance requirements. 

Understanding the characteristics of each sample matrix helps laboratories identify potential interferences and select appropriate corrective measures. 

Inadequate Quality Control Procedures 

Quality control procedures provide an essential means of evaluating whether laboratory mercury testing results are accurate, precise, and suitable for their intended purpose. Without appropriate quality control, analytical errors may remain undetected until results have already been reported. 

Several types of quality control samples help laboratories evaluate different aspects of analytical performance: 

  • Method blanks: Help identify contamination introduced through reagents, equipment, or laboratory preparation procedures. 
  • Laboratory duplicates: Evaluate analytical precision by comparing results from separately prepared or analyzed portions of the same sample. 
  • Matrix spikes: Assess the recovery of known mercury additions and help identify potential matrix-related interferences. 
  • Certified reference materials: Provide independently characterized mercury concentrations for evaluating analytical accuracy when suitable materials are available. 

Each quality control measure provides different information. A laboratory may obtain acceptable duplicate results while still experiencing systematic errors caused by contamination or incomplete mercury recovery. For this reason, relying on a single quality control procedure is insufficient. 

Laboratories should establish acceptance criteria based on the applicable analytical method, regulatory requirements, and validated laboratory procedures. Quality control results should be reviewed before analytical data are approved for reporting. 

When results fall outside established limits, laboratory personnel should investigate potential causes and document corrective actions. Depending on the issue, corrective measures may involve instrument maintenance, reagent replacement, additional sample preparation, or reanalysis. 

Maintaining detailed records of calibration checks, quality control results, and corrective actions also supports traceability and helps laboratories identify recurring analytical problems. 

Best Practices for Improving Mercury Testing Accuracy 

Improving mercury testing accuracy requires consistent attention to laboratory procedures throughout the analytical process. Even well-maintained instruments may produce unreliable results when samples are collected, prepared, or handled incorrectly. 

Standard operating procedures should clearly define requirements for sample collection, preservation, preparation, instrument operation, quality control, and reporting. These procedures should reflect the analytical methods used by the laboratory and be reviewed when methods or regulatory requirements change. 

Regular personnel training is equally important. Laboratory staff should understand the potential sources of mercury contamination, the consequences of improper sample handling, and the procedures used to identify analytical problems. 

Routine equipment maintenance and laboratory cleanliness can further reduce measurement errors. Instruments should be inspected and serviced according to manufacturer recommendations and laboratory requirements, while shared equipment should be cleaned using established procedures. 

Laboratories should also review quality control trends rather than evaluating each result in isolation. Gradual changes in blank concentrations, spike recoveries, or calibration verification results may indicate developing problems that require investigation. 

Finally, laboratories should periodically assess their testing procedures against applicable regulatory methods and established analytical standards. This includes confirming that analytical techniques, detection limits, preservation procedures, and quality control requirements remain appropriate for the samples being tested. 

Reliable laboratory mercury testing depends on controlling potential errors at every stage of the analytical process. Sample contamination, improper storage, preparation mistakes, instrument calibration problems, and matrix interferences can all affect reported concentrations. By maintaining consistent procedures, verifying analytical performance, and investigating quality control failures, laboratories can reduce measurement errors and improve the reliability of their results. These practices support defensible scientific findings, informed operational decisions, and compliance with applicable testing requirements. 

About the author 

Michael R. Halvorson, PhD, is Business Development Manager for AGS Scientific, a laboratory instrument distributor. Dr. Halvorson, who has more than 20 years of experience in the industry, focuses on biochemical toxicology, analytical methods development, sample preparation, laboratory automation and product development. 