Ultrapure Water System / Ultra Pure Water Systems

Ultra pure water is water nearly or completely devoid of contaminants. Ultrapure water systems are an important component for laboratories across markets including manufacturing and the semi-conductor industry. These ultra pure water systems remove chemical contaminants through de-ionization (DI) and reverse osmosis. An ultra pure water system eliminates bacteria with UV radiation. These laboratory water purification systems can achieve up to 18.2 MOhm-cm in resistivity (high resistivity indicating low levels of anionic contamination) and can reduce organic carbon to less than 50 ppb. An ultrapure water system is equipped with ultra filters to remove endotoxins, DNase and Rnase left over from bacteria destroyed by UV, resulting in extremely low total organic carbon (TOC). Additional carbon filters in these laboratory water purification systems reduces the need for pre-treatment, permitting tap water as a source.

Find, compare, and request a quote for ultrapure water systems across leading suppliers at Labcompare; including access to product citations, related published figures, and first-hand reviews from our scientific community.

CompanyPurite LtdPurite Ltd
ItemPurite HPAPurite Integra HP
Price
DescriptionCompact water purification system designed to guarantee a consistent supply of high quality pure water for clinical laboratory analyser feed. Integrates reverse osmosis, deionisation, and internal microfiltration with a colour LCD touch-screen display. Compatible with all major brands of clinical ... Read MoreCompact, centralised water purification and distribution unit for clinical analyser feed. Incorporates reverse osmosis and UV radiation technologies with internal storage and distribution pump. Produces Type I, II or III water, RO water, deionised water, or ultrapure water. Available in HP IT (... Read More
MethodReverse Osmosis (RO) + Deionisation (DI) + MicrofiltrationReverse osmosis (RO), UV (254 nm), deionisation (various polishing cylinder options), 5 µm pre-filter, optional 0.2 µm microfiltration, optional CO2 membrane degassing
Output30 L/hr @ 10 C; 48 L/hr @ 25 C (based on 4 bar feed water pressure)Purified water make-up: 120 or 190 L/hr (at 10°C); distribution: up to 270 L/hr at max 3 bar; minimum feed flow: 400 L/hr
Resistivity>10 MOhm/cm< 30 µS/cm to 18.2 MO·cm (with optional deionisation cylinder)
Dimensions440 mm (W) x 560 mm (D) x 750 mm (H)HP IT: 890 mm (W) x 500 mm (D) x 840 mm (H); HP GP: 1110 mm (W) x 604 mm (D) x 1842 mm (H)
Total Organic Carbon (TOC)<20 ppb< 30 ppb
Get Quote