| Description | The OmniaLabED+ is a large all-round ultrapure and pure water system designed for high-demand laboratories requiring both Type I ultrapure water and Type II pure water. Features continuously self-regenerating electro-deionization (EDI), a 100-liter recirculating storage tank with pressurized outlet,The OmniaLabED+ is a large all-round ultrapure and pure water system designed for high-demand laboratories requiring both Type I ultrapure water and Type II pure water. Features continuously self-regenerating electro-deionization (EDI), a 100-liter recirculating storage tank with pressurized outlet, and complies with ASTM, ISO 3696, and CLSI (CLRW) standards. Available in 20 L/h and 40 L/h variants with optional UV-TOC monitoring.... Read More | The OmniaLabDS is a CLRW/CLSI-grade pure water system designed for feeding clinical analyzers, steam sterilizers, and washer-disinfectors. Available in 20, 40, and 60 l/h output models, it features a 100 L pressurized tank with quality recirculation, regenerative polishing cartridges, a 0.2 µm The OmniaLabDS is a CLRW/CLSI-grade pure water system designed for feeding clinical analyzers, steam sterilizers, and washer-disinfectors. Available in 20, 40, and 60 l/h output models, it features a 100 L pressurized tank with quality recirculation, regenerative polishing cartridges, a 0.2 µm sterile filter, UV disinfection, and an optional emergency supply. Complies with DIN EN 285 and EN ISO 15883 international water standards.... Read More | Compact, area-efficient ultrapure water (Type I) system with OptiFill basic OneHand dispenser. Delivers 18.2 MO·cm resistivity and TOC <10 ppb (standard) or <5 ppb (UV variant). Integrated leakage sensor, pressure reducer, and pre-treatment cartridge. Bench or wall mountable. Available as standard (Compact, area-efficient ultrapure water (Type I) system with OptiFill basic OneHand dispenser. Delivers 18.2 MO·cm resistivity and TOC <10 ppb (standard) or <5 ppb (UV variant). Integrated leakage sensor, pressure reducer, and pre-treatment cartridge. Bench or wall mountable. Available as standard (art. 18000004) and UV (art. 18000005) models. Typical applications: AAS, IC, ICP, buffers and media preparation (standard); ultra-trace analysis, ICP-MS, HPLC, TOC-analysis (UV).... Read More | Compact pure water system with integrated reverse osmosis and deionization for Type II (ASTM II) pure water. Direct tap water connection, OptiFill touch dispenser standard, 6–20 L/h output, optional UV disinfection (254 nm), compatible storage tanks 10/30/60 L with recirculation. Applications: Compact pure water system with integrated reverse osmosis and deionization for Type II (ASTM II) pure water. Direct tap water connection, OptiFill touch dispenser standard, 6–20 L/h output, optional UV disinfection (254 nm), compatible storage tanks 10/30/60 L with recirculation. Applications: buffer and media preparation, feeding ultrapure water systems.... Read More | The OmniaTap xstouch (Blueline series) is an intelligent compact water purification system with intuitive touchscreen operation that simultaneously produces Type I ultrapure water and Type II pure water from a single unit connected directly to tap water. It combines RO, ion exchange, UV, TOC The OmniaTap xstouch (Blueline series) is an intelligent compact water purification system with intuitive touchscreen operation that simultaneously produces Type I ultrapure water and Type II pure water from a single unit connected directly to tap water. It combines RO, ion exchange, UV, TOC monitoring, and optional ultrafiltration. Available in two models: OmniaTap xstouch 8 UV-TOC and OmniaTap xstouch 8 UV-TOC/UF.... Read More |
| Method | Reverse Osmosis (RO) + Electro-Deionization (EDI) + UV | RO cartridge ? stainless steel regenerative polishing cartridge ? 0.2 µm sterile filter capsule ? UV disinfection; sterile overflow and sterile vent filter + CO2 absorber included; optional UV tank disinfection and emergency supply | Ion exchange with integrated pre-treatment cartridge; UV lamp (UV variant); recirculation function to prevent biofilm | Reverse osmosis (RO) + ion exchange deionization; optional UV disinfection at 254 nm; sterile filter capsule 0.2 µm (included) | Reverse osmosis (RO) + ion exchange + UV irradiation + TOC monitoring; UV-TOC/UF model additionally includes ultrafiltration (UF) for endotoxin, RNase, DNase, and protease removal |
| Dimensions | Tower: 511 x 1520 x 575 mm (W x H x D); Base cabinet (tank): 511 x 800 x 575 mm (W x H x D) | Tower: 511 x 1520 x 575 mm (W x H x D); Base cabinet (tank): 511 x 800 x 575 mm (W x H x D) | 240 x 530 x 435 mm (W x H x D) | Tower without tank: 390 x 720 x 525 mm (W x H x D); Tower with 10-liter docking tank: 390 x 720 x 615 mm (W x H x D) | Without tank (with OptiFill touch dispenser): 240 x 530 x 435 mm (W x H x D); With 7 L docking tank: 253 x 530 x 520 mm; Production unit without tank (under-table): 240 x 415 x 310 mm; OptiFill touch wall dispenser: 90 x 482 x 337 mm; OptiFill touch bench dispenser: 250 x 650 (750 with extension) x 380 mm |
| Total Organic Carbon (TOC) | < 10 ppb (standard models); < 5 ppb (UV-TOC monitoring models) | < 2 ppm (feed water requirement) | <10 ppb (standard); <5 ppb (UV variant) | < 30 ppb (typical; varies with feed water quality) | < 5 ppb (Type I ultrapure water, both models); real-time TOC monitoring included |