Spectrophotometers are used to measure the concentration of a known substance in a solution. They do this by passing a light through the substance and measuring light intensity as a function of wavelength. Either single beam or double beam spectrophotometers can be purchased. In a single beam spectrophotometer, a reference standard is used to measure light intensity before and after the sample is loaded. For double beam spectrophotometers the beam is split with one beam passing through the standard and the other through the sample to compare intensities. While single beam spectrophotometers are typically less expensive and have higher sensitivity, double beam spectrophotometers tend to be more stable and easier to use. Spectrophotometers are important pieces of equipment in chemistry, physics, biochemistry, clinical, and materials testing labs.
Find, compare, and request a quote for spectrophotometers across leading suppliers at Labcompare; including access to product citations, related published figures, and first-hand reviews from our scientific community.
|  |  |
| Company | Agilent Technologies | Agilent Technologies |
| Item | Cary 60 UV-Vis Spectrophotometer | Cary 5000 UV-Vis-NIR Spectrophotometer |
| Citations | | |
| Figures | | |
| Reviews | | |
| Catalog Number | Cary 60 | 10079300 |
| Price | | |
| Type | UV-Visible | UV-VIS-NIR |
| Wavelength Range | 190 to 1100 nm | 175 to 3300 nm |
| Wavelength Accuracy | ± 0.5 at 541.94 nm | Inquire |
| Scanning Speed | 24,000 nm/min, 80 data points/sec | UV-Vis 2000 nm/min; NIR 8000 nm/min |
| Spectral Bandwidth | 1.5 nm fixed | Down to 0.01 nm |
| Optical System | Double beam | Double Beam |
| Light Source | Xenon flash lamp | Tungsten halogen visible and deuterium arc UV |
| Resolution | Inquire | Inquire |
| Description | The Cary 60 UV-Vis spectrophotometer has a wavelength range of 190–1100 nm that can be scanned in under three seconds. Our flexible UV-Vis system can be fitted with long pathlength cuvettes and solid sample transmission or reflectance accessories, and is ideal for remote UV-Vis absorbance analysis, The Cary 60 UV-Vis spectrophotometer has a wavelength range of 190–1100 nm that can be scanned in under three seconds. Our flexible UV-Vis system can be fitted with long pathlength cuvettes and solid sample transmission or reflectance accessories, and is ideal for remote UV-Vis absorbance analysis, when fitted with fiber optic probes.
The xenon source lamp has a 10 year replacement warranty (for Cary 60 instruments purchased from Agilent or participating partners), requires zero warmup, and causes no photodegradation of samples. Fast reactions can be recorded with 80 data points per second. The Cary 60 is well-suited as a routine UV-Vis spectrophotometer or for use in teaching labs. Optional 21 CFR Part 11 software tools are available.
Features - Reduce time-consuming and expensive lamp replacements, and minimize instrument revalidation costs with the unique Agilent xenon source lamp.
- Safely measure life-science samples without sample photodegradation and ensure you get the correct answer every time with the Cary 60 UV-Vis spectrophotometer.
- Get reliable UV-Vis spectrophotometer performance with fiber-optic measurement probes. Why use cuvettes when you can bring the UV-Vis instrument to the sample and get accurate results in a fraction of the time?
- Measure small volumes and precious samples accurately and reproducibly with a highly focused beam image.
- Scan the entire wavelength range (190 to 1100 nm) in under three seconds and collect data from single or multiple wavelengths at 80 data points per second.
- Take advantage of the Cary 60 UV-Vis spectrophotometer’s comprehensive range of accessories to accurately characterize the widest variety of liquid and solid samples.
- Tailor the Cary WinUV software to suit your analytical requirements. From classroom to research laboratories through R&D and regulated QA/QC environments, simplify your UV-Vis spectroscopy measurements and finish more in less time.
- Choose sustainability: The Cary 60 UV-Vis received My Green Lab’s ACT (Accountability, Consistency, Transparency) label after an independent audit for its environmental impact throughout the product lifecycle.
... Read More | The Cary 5000 is a high-performance UV-Vis and NIR spectrophotometer with superb photometric performance in the 175–3300 nm range. Using a PbSmart detector, the Cary 5000 extends its NIR range to 3300 nm making it a powerful tool for materials science research. The Cary WinUV software, a modular The Cary 5000 is a high-performance UV-Vis and NIR spectrophotometer with superb photometric performance in the 175–3300 nm range. Using a PbSmart detector, the Cary 5000 extends its NIR range to 3300 nm making it a powerful tool for materials science research. The Cary WinUV software, a modular Windows-based software, makes it easy to perform powerful analysis and control a number of optional accessories.
The large sample compartment can be expanded to hold large accessories and integrating spheres for spectral and diffuse reflectance. The LockDown mechanism makes it possible to quickly change and position accessories for reproducible results.
Features - Measure beyond 8.0 absorbance units with reference beam attenuation
- Measure 175 to 3300 nm using a PbSmart NIR detector for extended photometric range
- WinUV software—modular software with power analysis and enhanced transfer and report export capabilities
- Versatile set of accessories for materials research including spectral and diffuse reflectance
- Variable slit widths (down to 0.01 nm) for optimum control over data resolution
- Large sample compartment with LockDown mechanism for quick changes and positioning for reproducible results
- Maximum light throughput using Schwarzchild coupling optics for higher accuracy at low transmission levels
- Minimal noise and stray light using a floating aluminum casting and double Littrow monochromator
- Extended dynamic range by attenuating the reference beam to balance double beam signal levels.
- Nitrogen purging—separate purging in the sample and monochromator compartments to reduce oxygen and water vapor in the ultra-violet.
... Read More |
| Quantity | EA | EA |
| Get Quote | | |